5-Part Auto Hematology Analyzer Hemaray 86 for sale
MODEL / Hemaray 865-Part Auto Hematology Analyzer Hemaray 86 for sale. Integral Titanium Incubation System
- 28 Parameters
- 2 Histograms, 2 Scattergrams
- 60 samples/hour
- Autoloader for 50 sample tubes
Product Details
Quick Details
28 Parameters
2 Histograms, 2 Scattergrams
60 samples/hour
Autoloader for 50 sample tubes
3-D Topographic Maps for WBC
Integral Titanium Incubation System
Laser Scatter Technology
Constant Linear Sheath Flow
Random Direction Tube Barcode Scanning Technology (Optional)
Specifications
Hemaray 86 Auto Hematology Analyzer
* Features
28 Parameters
2 Histograms, 2 Scattergrams
60 samples/hour
Autoloader for 50 sample tubes
3-D Topographic Maps for WBC
Integral Titanium Incubation System
Laser Scatter Technology
Constant Linear Sheath Flow
Random Direction Tube Barcode Scanning Technology (Optional)

* Technical Specifications
* Technology
3-D Topographic Map
Integral Titanium Incubation System
Laser Scatter Technology
Flow Cytometry Technology
Cyanide Free Reagents
Random Direction Tube Barcode Scanning Technology
* Parameters
Regular Parameters
WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-SD,RDW-CV, PLT, PDW, MPV, PCT, P-LCR, NEU#,NEU%,LYM#,LYM%,MON#,MON% EOS# EOS%, BAS#, BAS%
Research Parameters
AL Y#, ALY%, IG#, IG%
2 Histograms, 2 Scattergrams
* Test Speed
Autoloader Mode: 60 Samples/hour
Close Tube Mode: 60 Samples/hour
* Test Mode
CBC+5DIFF, CBC
* Sampling Mode
Autoloader: 50 tubes position random direction tube barcode scanning technology
Close Tube: STAT priority, support both whole blood and capillary blood samples
* Performance

Item Linear range CarryOver CV
WBC 1.0-99.9x109/L ≤1.0% ≤2.0%
RBC 0.3-7.0x1012/L ≤1.0% ≤1.5%
HGB 20-240g/l ≤1.0% ≤1.5%
PLT 20-999x109/L ≤1.0% ≤4.0%
* QC and Calibration
Multiple QC rules, including L-J, X-B etc
Automated and manual calibration function for both whole blood mode and pre-diluted blood mode
* Power Supply
100V-240V, 50Hz/60Hz

* Operation Environment
Temperature: 15℃-30℃
Humidity: 30%-85%
* Weight: 75KG
* Dimension: 61cm (L) ×76cm (W) ×61cm (H)
Technical Specifications
Technical parameters from the currently available product information.
- Flow Cytometry Technology
- Cyanide Free Reagents
- Random Direction Tube Barcode Scanning Technology
- * Parameters
- Autoloader Mode
- 60 Samples/hour
- Close Tube Mode
- 60 Samples/hour
- * Test Mode
- CBC+5DIFF, CBC
- Autoloader
- 50 tubes position random direction tube barcode scanning technology
- Close Tube
- STAT priority, support both whole blood and capillary blood samples
- Item Linear range CarryOver CV
- WBC 1.0-99.9x109/L ≤1.0% ≤2.0%
- * Power Supply
- 100V-240V, 50Hz/60Hz
- Temperature
- 15℃-30℃
- Humidity
- 30%-85%
- * Weight
- 75KG
- * Dimension
- 61cm (L) ×76cm (W) ×61cm (H)
Specifications can vary by configuration. Confirm intended use, final model, and supply details with our product team.