Surgery

Cost effective portable Diabetes Insulin Infusion Pump EKONIS32

MODEL / EKONIS32

Insulin pump is a portable device which delivers a specific amount of insulin in the body at specific time interval.

  • Insulin pump is a portable device which delivers a specific amount of insulin in the body at specific time interval.
  • CPU: dual CPU
  • Store condition: pls store it in ventilated rooms without any corrosive gases, temperature is from -20℃~55℃, Relative humidity no more than 80%.
  • IP grade: IPX8
PRODUCT OVERVIEW

Product Details

Quick Details

Insulin pump is a portable device which delivers a specific amount of insulin in the body at specific time interval.

Specifications

1. Size: 79*52*20mm 2. Weight: 58g 3. Basal rate segment: 24/48 segments for optional. 4. Infusion frequency: 0--60 times/hour 5. Minimal dose: 0.05u (1u means 0.01ml) 6. Infusion type: basal rate, Temporary basal rate, Bolus, Square wave bolus. 7. Basal rate setting range: 0--5u 8. Basal rate infusion error range: +-5% 9. Temporary basal rate setting range: speed: 0--200%, time: 0--24 hours. 10. Bolus setting range: normally 0--40u, max setting range: 0--85u

Cost effective portable Diabetes Insulin Infusion Pump EKONIS32

11. Bolus infusion error range: +-3% 12. Bolus modes: normal bolus, bolus before 3 meals. 13. Square wave bolus setting range: normally 2--40u, max setting range: 2--85u, time:0--240 minutes. 14. Max daily dose: 200u 15. Screw rewind: motor drive rewind automatically 16. Alarm: ring 17. Alarm situation: A: Meal alarm: 30 minutes after bolus infusion B: Low-Voltage alarm: voltage lower than 1.2v C: No liquid alarm: liquid less than 20u. D: No infusion alarm: no liquid left. E: Liquid blocking alarm: blocking before 6u

Cost effective portable Diabetes Insulin Infusion Pump EKONIS32

18. CPU: dual CPU 19. Reservoir volume: 320u 20. Infusion tube interface: luer interface 21. Battery type: 1*AAA battery 22. Using condition: Temperature: +5℃~+40℃, Relative humidity: 20%~90%, Atmospheric pressure: 700hPa~1060hPa. 23. Store condition: pls store it in ventilated rooms without any corrosive gases, temperature is from -20℃~55℃, Relative humidity no more than 80%. 24. IP grade: IPX8

TECHNICAL DATA

Technical Specifications

Technical parameters from the currently available product information.

Basal rate segment
24/48 segments for optional.
Infusion frequency
0--60 times/hour
Minimal dose
0.05u (1u means 0.01ml)
Infusion type
basal rate, Temporary basal rate, Bolus, Square wave bolus.
Basal rate setting range
0--5u
Basal rate infusion error range
+-5%
Temporary basal rate setting range
speed: 0--200%, time: 0--24 hours.
Bolus setting range
normally 0--40u, max setting range: 0--85u
Bolus infusion error range
+-3%
Bolus modes
normal bolus, bolus before 3 meals.
Square wave bolus setting range
normally 2--40u, max setting range: 2--85u, time:0--240 minutes.
Max daily dose
200u
Screw rewind
motor drive rewind automatically
B: Low-Voltage alarm: voltage lower than
1.2v
Reservoir volume
320u
Infusion tube interface
luer interface
Battery type
1*AAA battery
Using condition
Temperature: +5℃~+40℃, Relative humidity: 20%~90%, Atmospheric pressure: 700hPa~1060hPa.