Other

EKONBI18 Infant Radiant Warmer | Neonatal Care

MODEL / EKONBI18

Main body (including the warmer module, temperature controller, bassinet, stand), I.V. Pole, tray.

  • Set Temperature and Baby Temperature can be displayed separately
  • Horizontal angle of warmer module and inclination of the bassinet can be adjusted
  • The panels surrounded bassinet can be disassembled
  • Multiple failure alarm indications
PRODUCT OVERVIEW

Product Details

Quick Details

Brief Introduction

● Three control Modes: Pre-warm Mode, Manual Mode and Baby Mode controlled by micro-computer;

● Set Temperature and Baby Temperature can be displayed separately;

● Horizontal angle of warmer module and inclination of the bassinet can be adjusted;

● The panels surrounded bassinet can be disassembled;

● APGAR timer;

● Multiple failure alarm indications;

● RS232 connector for data output;

● Disposable skin temperature sensor is optional

Specifications

EKONBI18 Infant Radiant Warmer | Neonatal Care

Brief Introduction

● Three control Modes: Pre-warm Mode, Manual Mode and Baby Mode controlled by micro-computer;

● Set Temperature and Baby Temperature can be displayed separately;

● Horizontal angle of warmer module and inclination of the bassinet can be adjusted;

● The panels surrounded bassinet can be disassembled;

● APGAR timer;

● Multiple failure alarm indications;

● RS232 connector for data output;

● Disposable skin temperature sensor is optional

neonatal incubator

EKONBI18 Infant Radiant Warmer | Neonatal Care

Specification

Power requirement: AC220V-230V/50Hz or AC110-120V/50-60Hz, 600VA

Control Mode: Pre-warm Mode, Manual Mode and Baby Mode control by micro-computer

Baby Mode temperature control range: 34.5-37.5℃

Skin temperature display range: 5℃-65℃

Accuracy of skin temperature sensor: ±0.2℃

Temperature uniformity of mattress: ≤2℃

Angle of heater head: 0°, 30°, 45°for two ways

Inclination of the bassinet: five tilt position

APGAR timer: Audible and visible alarm when the device runs to 50″~1′、4′50″~5′、9′50″~10′

Failure alarm: over temperature alarm, deviation alarm, sensor failure alarm, power failure alarm, setting alarm, checking alarm and so on.

Standard Configuration

Main body (including the warmer module, temperature controller, bassinet, stand), I.V. Pole, tray.

Environmental Requirement

Operating range: 18 ℃~30℃

Ambient air velocity rate: <0.3m/s

MSL TEAM picture

Welcome to medicalequipment-msl.com.

TECHNICAL DATA

Technical Specifications

Technical parameters from the currently available product information.

Three control Modes
Pre-warm Mode, Manual Mode and Baby Mode controlled by micro-computer
Power requirement
AC220V-230V/50Hz or AC110-120V/50-60Hz, 600VA
Control Mode
Pre-warm Mode, Manual Mode and Baby Mode control by micro-computer
Baby Mode temperature control range
34.5-37.5℃
Skin temperature display range
5℃-65℃
Accuracy of skin temperature sensor
±0.2℃
Temperature uniformity of mattress
≤2℃
Angle of heater head
0°, 30°, 45°for two ways
APGAR timer
Audible and visible alarm when the device runs to 50″~1′、4′50″~5′、9′50″~10′
Failure alarm
over temperature alarm, deviation alarm, sensor failure alarm, power failure alarm, setting alarm, checking alarm and so on.
Operating range
18 ℃~30℃
Ambient air velocity rate
<0.3m/s

Specifications can vary by configuration. Confirm intended use, final model, and supply details with our product team.