EKONMD23S Portable Ultrasonic A/B Scanner for Ophthalmology
MODEL / EKONMD23SThis product is intended for ophthalmological ultrasonic diagnosis and AL biometric measurement.
- EKONMD23S Portable Ultrasonic A/B Scanner Software Version
- EKONMD23S Portable Ultrasonic A/B Scanner Expected Service Life
- EKONMD23S Portable Ultrasonic A/B Scanner Basic Working Principle
- EKONMD23S Portable Ultrasonic A/B Scanner Working Conditions
Product Details
Quick Details
This product is intended for ophthalmological ultrasonic diagnosis and AL biometric measurement.
Specifications
EKONMD23S Portable Ultrasonic A/B Scanner for Ophthalmology

EKONMD23S Portable Ultrasonic A/B Scanner Introduction
The Ultrasonic A/B Scanner for Ophthalmology is an ultrasonic imaging instrument specialized in ophthalmological diagnosis. It consists of a main unit, a power adaptor, a 10MHz B-Probe, a 20MHz B-Probe, a 10MHz A-Probe, a footswitch, a keyboard and a mouse.
The 10MHz B-Probe applies to normal ophthalmological ultrasonic diagnosis; and the 20MHz B-Probe applies to observation of details of intraocular and retina structure.
Patients with eyelid trauma and severe eye infection are prohibited from using B-Scan, and patients with keratitis and cornea trauma are prohibited from using A-Biometric scan.
EKONMD23S Portable Ultrasonic A/B Scanner Software Version
MD24S V1.0.

EKONMD23S Portable Ultrasonic A/B Scanner Expected Service Life
Based on the experience of products sold (and considering technology upgrade cycle), as well as the aging of major parts - transducer which will probably reduce the basic performance, the product's expected service life is determined as five years according to the normal usage of six hours each working day. (Service life may extend if the major parts, like the probe, are returned to the manufacturer for upgrade.)
EKONMD23S Portable Ultrasonic A/B Scanner Basic Working Principle
a) B-Mode: Under the excitation of electric pulse, the 10MHz/20MHz ultrasound transducer emits ultrasonic pulse into eye and receives reflective echo from the surfaces of different tissue, thus forming real-time two-dimensional image of eye tissue.
b) A-Mode: Under the excitation of electric pulse, the 10MHz ultrasound transducer emits ultrasonic pulse into eye and receives reflective echo from the surfaces of three different tissues of anterior chamber, lens and vitreous body. The transmitting time of ultrasonic wave in different tissues are measured and the distance of each part will be calculated according to the acoustic velocity of different tissues and then the axial length (AL) will be calculated.

EKONMD23S Portable Ultrasonic A/B Scanner Working Conditions
Adaptor
1. Working Temperature: 5℃-40℃;
2. Relative Humidity: lower than 80%;
3. Input Voltage of Power Adaptor: AC 100-240V, 50/60Hz;
4. Atmospheric Pressure: 70-106kPa.
Main Unit
1. Working Temperature: 5℃-40℃;
2. Relative Humidity: lower than 80%;
3. Input Voltage of Main Unit: DC 12V, 4A
4. Atmospheric Pressure: 70-106kPa.
Basic Parameters
1. Nominal Ultrasound Frequency: B-Probe: 10MHz/20MHz Available, A-Probe: 10MHz;
2. User Interface: monitor, keyboard, mouse, touch screen and footswitch;
3. Monitor: resolution≥1024×768, gray scale≥256;
4. Keyboard and Mouse: USB interface, wireless keyboard and mouse; see Annex D for the main specification;
5. Output Interface: 4×USB2.0 and HDMI interface;
6. Net Weight: 5.5kg;
7. Size: 34cm x 31cm x8cm.
EKONMD23S Portable Ultrasonic A/B Scanner Specifications
B-mode
1. Scanning Method: mechanical sector scan;
2. Display Mode: B, B+A;
3. Scanning Angle: 53;
Display Depth Adjustable Range: 10MHz: 28mm~60mm; 20MHz: 19 mm~40mm;
4. Variable Delayed Depth: 10MHz: 0 mm~15 mm; 20MHz: 0 mm~10mm;
5. Gain Control Range: 1-105dB;
6. TGC: -20dB~20dB with dynamic range, manual sectional adjustment (6-section adjustable);
7. Grey Scale: 256 Levels;
8. Frame Rate: 10fps;
9. Image Saving: 100 images;
10. Dynamic Loop: 10s/100 images, loop or frame-by-frame playback;
11. Image Processing and Signal Post-processing: Frame-Averaging, Pseudo-Color Codes, and Gamma Correction
12. Axial Resolution
a) 10MHz Probe: No more than 0.1mm;
b) 20MHz Probe: No more than 0.08mm;
13. Lateral Resolution
a) 10MHz Probe: No more than 0.2mm;
b) 20MHz Probe: No more than 0.15mm.
A-mode
1. AL Biometric Measuring Range: Axial Length (AL): 15mm-40mm;
2. AL Biometric Measuring Accuracy: No more than ±0.05mm;
3. Total Gain: 98dB, user’s gain adjustability: 1~60dB;
4. Measuring Mode: 5 groups (Normal, Aphakic, Special, Cataract and Manual Measurement);
6. IOL calculation: SRK-T, SRK-II, BINK-II, HOLLADAY, HOFFER-Q and HAIGIS;
7. Support IOL calculation after refractive surgery, two groups of formulas can be calculated contrastively and displayed simultaneously.
8. Automatic measurement and 10 groups averaging and display standard deviation.
9. Up to 8 groups of IOL constants can be stored.
Safety
1. The electrical safety of this product shall be in accordance with requirements of IEC60601-1, IEC60601-1-2 and IEC60601-2-37.
2. For information on Acoustic Output Parameters, please see Annex B.
MSL TEAM picture
Welcome to medicalequipment-msl.com,If you have any demand in ultrasound machine.
Technical Specifications
Technical parameters from the currently available product information.
- a) B-Mode
- Under the excitation of electric pulse, the 10MHz/20MHz ultrasound transducer emits ultrasonic pulse into eye and receives reflective echo from the surfaces of different tissue, thus forming real-time two-dimensional image of eye tissue.
- b) A-Mode
- Under the excitation of electric pulse, the 10MHz ultrasound transducer emits ultrasonic pulse into eye and receives reflective echo from the surfaces of three different tissues of anterior chamber, lens and vitreous body. The transmitting time of ultrasonic wave in different tissues are measured and the distance of each part will be calculated according to the acoustic velocity of different tissues and then the axial length (AL) will be calculated.
- Working Temperature
- 5℃-40℃
- Relative Humidity
- lower than 80%
- Input Voltage of Power Adaptor
- AC 100-240V, 50/60Hz
- Atmospheric Pressure
- 70-106kPa.
- Input Voltage of Main Unit
- DC 12V, 4A
- Nominal Ultrasound Frequency
- B-Probe: 10MHz/20MHz Available, A-Probe: 10MHz
- User Interface
- monitor, keyboard, mouse, touch screen and footswitch
- Monitor
- resolution≥1024×768, gray scale≥256
- Keyboard and Mouse
- USB interface, wireless keyboard and mouse; see Annex D for the main specification
- Output Interface
- 4×USB2.0 and HDMI interface
- Net Weight
- 5.5kg
- Display Mode
- B, B+A
- Scanning Angle
- 53?
- Display Depth Adjustable Range
- 10MHz: 28mm~60mm; 20MHz: 19 mm~40mm
- Variable Delayed Depth
- 10MHz: 0 mm~15 mm; 20MHz: 0 mm~10mm
- Gain Control Range
- 1-105dB
Specifications can vary by configuration. Confirm intended use, final model, and supply details with our product team.
Discuss this product with our team
Share your application, destination, and required configuration for a project-specific response.