Fluorescence Immunoassay Chromatography Analyzer EKONS100

Product Performance

Accuracy: Detect WTO CRP standard material (NIBSC code:85/506), error should be within ±15%.

Measurement Precision: Detect fluorescent latex microspheres [F1 -XC(39376)] of Merck Chemicals Company, coefficient of variation (CV) shall be within 15%.

Stability of light detection: Detect fluorescent latex microspheres [F1 -XC(39376)] of Merck Chemicals Company, and detect 10 times continuous, test time interval 30mins, RHL≤3%.

Linear Range: In 2 order of magnitude range, the linear dependent coefficient (r)≥0.9900.

quantitative measurement

Fluorescence Immunoassay Chromatography Analyzer EKONS100

Scope of application

The product is intended use for immune test on fluorescent immune chromatography method, combined used with POCT test kits manufactured by the company.

The analyzer is widely used in medical institutions of central laboratory, out patient /emergency laboratory, clinical departments, and other medical services (such as community clinic), physical examination center, corollary using with fluorescent immune chromatography test kits can be used for patients’ early screening, or health check-up.

quantitative measurement

Fluorescence Immunoassay Chromatography Analyzer EKONS100

Working Condition

Power supply: AC220V 50Hz;

Input Power: 100VA;

Environment Temperature (5~40°C)

Relative Humidity 10%~85%

Atmospheric Pressure 86.0kPa~106.0kPa

quantitative measurement
quantitative measurement

MSL TEAM picture

Welcome to medicalequipment-msl.com.

TECHNICAL DATA

Technical Specifications

Technical parameters from the currently available product information.

Accuracy
Detect WTO CRP standard material (NIBSC code:85/506), error should be within ±15%.
Measurement Precision
Detect fluorescent latex microspheres of Merck Chemicals Company, coefficient of variation (CV) shall be within 15%.
Stability of light detection
Detect fluorescent latex microspheres of Merck Chemicals Company, and detect 10 times continuous, test time interval 30mins, RHL≤3%.
Linear Range
In 2 order of magnitude range, the linear dependent coefficient (r)≥0.9900.
Power supply
AC220V 50Hz;
Input Power
100VA;