Full Automatic Endoscope Irrigation Pump EKONGY01
MODEL / EKONGY01Endoscope Irrigation Pump EKONGY01 is suitable for Arthroscopy, Hysteroscopy and Urology systems etc.
- Outstanding flow pressure sensor, high sensitivity, very safe and stable
- Used in orthopaedics,Gynecology and Urology systems.
- Use roller pump to get large flow,irrigation tube can be autoclavable sterilization to avoid secondary pollution.
- Sterile tubing sets with sensor,a peristaltic pump,with two channels.
Product Details
Quick Details
Endoscope Irrigation Pump EKONGY01 is suitable for Arthroscopy, Hysteroscopy and Urology systems etc.
Pressure presetting range: 15-600mmHg(2-80kPa)
Flow rate: 0.1-1.0l/min
Sterile tubing sets with sensor, a peristaltic pump,with two channels
Emulsion tube and shim are replaceable (5pcs, L: approx.16cm)
Outstanding flow pressure sensor, high sensitivity, very safe and stable
Specifications
Full Automatic Endoscope Irrigation Pump EKONGY01
Endoscope Irrigation Pump EKONGY01 is suitable for Arthroscopy, Hysteroscopy and Urology systems etc.
Pressure presetting range: 15-600mmHg(2-80kPa)
Flow rate: 0.1-1.0l/min
Sterile tubing sets with sensor, a peristaltic pump,with two channels
Emulsion tube and shim are replaceable (5pcs, L: approx.16cm)
Outstanding flow pressure sensor, high sensitivity, very safe and stable

Full Automatic Endoscope Irrigation Pump EKONGY01
Main Characteristic:
1. Used in orthopaedics,Gynecology and Urology systems.
1. Use roller pump to get large flow,irrigation tube can be autoclavable sterilization to avoid secondary pollution.
2. Sterile tubing sets with sensor,a peristaltic pump,with two channels.
3. Emulsion tube and shim are replaceable(5pce,L:approx.16cm).
4. High sensitivity pressure sensor.
5. Pressure is adjustable on request and real time monitored.

Full Automatic Endoscope Irrigation Pump EKONGY01
Technical parameters:
Full Automatic Endoscope Irrigation Pump EKONGY01
TECHNICAL DATA
Power source: AC110\220V±10%50/60HZ 0.5A
Fuses: φ5×20mm250V 2A
Max.power consumption: 100W
Pressure presetting range: 15-600mmHg(2-80kPa,error approx.+2kPa)
Flow rate: 0.1-1.0l/min (error approx.+20%)
Signal/Noise ratio: ≤55dB
Safety standard: IEC601-1
Storage temp.: -15℃ to +50℃
Operating temp.: 0℃ to 45℃
Relative humidity:30% to 85%
N.W.: 8kg
Dims: 385x385x120mm
STANDARD FEATURES
Main unit
·Sterile tubing sets with sensor
·Shims of sensor (3pcs), emulsion tube and shim (5pcs, L: approx.16cm)
·Power cable

MSL TEAM picture
Welcome to medicalequipment-msl.com.
Technical Specifications
Technical parameters from the currently available product information.
| Supply voltage | ~220V 50Hz |
| Power | 150VA |
| Range of press | 15~400mmHg(2~53.3kPa) |
| Range of flw | 0.1~1.0L/min |
| Noise | ≤70dB |
Specifications can vary by configuration. Confirm intended use, final model, and supply details with our product team.
Discuss this product with our team
Share your application, destination, and required configuration for a project-specific response.