Low Sample Consumption Auto Hematology Analyzer EKONAB24
MODEL / EKONAB24EKON Medical cooperate with DHL,FEDEX,UPS,EMS,TNT,etc.International shipping company,make your goods? arrive destination safely and quickly.
- Compact, powerful and affordable
- LAN port supports HL7 protocol
- EKON Medical cooperate with DHL,FEDEX,UPS,EMS,TNT,etc.International shipping company,make your goods? arrive destination safely and quickly.
Product Details
Quick Details
Principles
Impedance method for WBC, RBC, PLT counting
Cyanide-free reagent for Hemoglobin test
Flow Cytometry (FCM), Semi-conductor Laser scatter, chemical dye method, independent Basophil channel
Specifications
Low Sample Consumption Auto Hematology Analyzer EKONAB24
Auto Hematology Analyzer EKONAB24 Features:
5-part differentiation, 29 parameters, 1 3D scatter gram, 3 2D scatter grams and 3 histograms
Tri-angle semi-conductor Laser scatter combined with chemical dye method, advanced flow cytometry
Compact, powerful and affordable
Only 20ul sampling volume
Up to 80 samples per hour
3 counting modes: venous blood, capillary blood and prediluted
Capability to flag abnormal samples
Large storage capacity: up to 100,000 samples

Auto Hematology Analyzer EKONAB24 Technical Specifications:
Principles
Impedance method for WBC, RBC, PLT counting
Cyanide-free reagent for Hemoglobin test
Flow Cytometry (FCM), Semi-conductor Laser scatter, chemical dye method, independent Basophil channel
Parameters
WBC, Lym%, Lym#,Mon%, Mon#,Neu%, Neu#,Eos%, Eos#, Bas%, Bas#, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, PDW, MPV, PCT,and 4 research parameters P-LCC, P-LCR, LIC%, LIC#, ALY%, ALY#
1 3D and 3 2D scatter grams and 3 histograms
Throughput
Up to 80 samples per hour
Sample Volume
Whole Blood: 20ul
Prediluted: 20ul
Test Mode
CBC CBC+DIFF
Performance
Carryover Repeatability Linearity
WBC ≤0.5% ≤2.0% (4-15×10 9 /L) 0.00-300×10 9 /L
RBC ≤0.5% ≤1.5% (3.5-6.0×10 12 /L) 0.00-8.5×10 12 /L
HGB ≤0.5% ≤1.5% (110-180g/L) 0-250g/L
PLT ≤1.0% ≤4.0% (150-500×10 9 /L) 0.00-3000×10 9 /L
Data Storage Capacity
Up to 100,000 results including numeric and graphical information
Communication
LAN port supports HL7 protocol
Operating Environment
Temperature: 15 o C-30 o C
Humidity: 30-85%
Air Pressure: 70-106 kPa
Power Requirement
A.C. 100-240V, ≤300VA, 50/60Hz
Dimension and Weight
620mm(L) × 620mm(W) × 535mm(H)
Weight: 56kg

MSL TEAM picture
Welcome to medicalequipment-msl.com,If you have any demand in ultrasound machine.
Technical Specifications
Technical parameters from the currently available product information.
- counting modes
- venous blood, capillary blood and prediluted
- Large storage capacity
- up to 100,000 samples
- Whole Blood
- 20ul
- Prediluted
- 20ul
- Temperature
- 15oC-30oC
- Humidity
- 30-85%
- Air Pressure
- 70-106 kPa
Specifications can vary by configuration. Confirm intended use, final model, and supply details with our product team.
Discuss this product with our team