Lab Equipment > Mindray analyzer

Mindray BS 120 fully blood chemistry analyzer price

MODEL / Mindray BS 120

Automatic, discrete, random access; Automatic probe cleaning, liquid level detection & collision protection; Automatic dilution for abnormal sample.

  • Automatic, discrete, random access
  • Automatic probe cleaning, liquid level detection & collision protection
  • Automatic dilution for abnormal sample
  • Automatic sample dilution
PRODUCT OVERVIEW

Product Details

Quick Details

Automatic, discrete, random access,

STAT sample priority

Specifications

Mindray BS 120 fully blood chemistry analyzer

Chemistry Analyzer

Discrete, random access, fully automated

100 tests per hour, up to 300 tests per hour with ISE

Up to 33 onboard chemistries and 3 ions

Refrigerated reagent compartment

Flexible configuration for sample/reagent positions

Automatic probe cleaning, liquid level detection & collision protection

8 wavelengths: 340~670nm

Automatic dilution for abnormal sample

External bar code reader (optional)

Bi-directional LIS interface

Mindray BS 120

Mindray BS 120 fully blood chemistry analyzer

Chemistry Analyzer

Technical Specifications

System Function:

Automatic, discrete, random access,

STAT sample priority

Throughput: Up to 100 tests/hour (without ISE), up to

300 tests/hour with ISE (3 ions)

Measuring principles:

Absorbance photometry, turbidimetry,

Ion Selective Electrode technology

Methodology: End-point, fxed-time, kinetic, optional

ISE

Single/dual reagent chemistries,

monochromatic/bichromatic

linear/non-linear multipoint calibration

Programming: Open system with user defned profles

and calculations

Reagent/Sample Handling:

Reagent/Sample tray:

Up to 33 positions for sample,

up to 35 positions for reagent;

24 hour non-stop refrigerated

compartment (4~15?C)

Reagent volume:

R1: 180~450μl, step by 1μl

R2: 30~250μl, step by 1μl

Sample volume: 3~45μl, step by 0.5μl

Reagent/Sample probe:

Liquid level detection, collision

protection and inventory checking

Probe cleaning: Automatic washing both interior and exterior

Carry-over < 0.1%

Automatic sample dilution:

Pre-dilution and post-dilution

dilution ratio up to 1: 150

Dilution vessel: Disposable cuvette

External Bar Code Reader (optional):

Used for sample and reagent programming;

Applicable to various bar code systems

including Codabar, ITF (Interleaved Two of

Five), Code128, Code39, UPC/EAN, Code93;

capable to communicate with LIS in a

bi-directional mode

mindray chemistry analyzer

ISE Module (optional):

Measure parameter: K +, Na +, Cl?

Reaction System:

Reaction rotor: Rotating tray, containing 40 cuvettes

Cuvette: Optical length 5mm

Reaction volume: 180~500μl

Reaction temperature: 37±0.1?C

Mixing system: Independent mixing bar

Optical System:

Light Source: Halogen-tungsten lamp

Wavelength: 340nm, 405nm, 450nm, 510nm,

546nm, 578nm, 630nm, 670nm

Linear range: 0~3.5Abs

Control and Calibration:

Calibration mode: Linear (one-point, two-point and multi-point),

Logit-Log 4P, Logit-Log 5P, Spline,

Exponential, Polynomial, Parabola

Control rules: Westgard multi-rule, Cumulative

sum check, Twin plot

Operation Unit:

Operation system: Windows? XP Professional/Home SP2, Windows? 7

or above Windows? VISTA Home/Business

Interface: RS-232

Working Conditions:

Power Supply: AC 200~240V, 50/60Hz, 800W or

AC 100~130V, 50/60Hz, 800W

Temperature: 15~30?C

Humidity: 35~85%

Water consumption: 2.5L/hour

Dimension: Bench top: 690mm(W)x570 mm(D)x595 mm(H)

Weight: 75 Kg

biochemistry analyzer

MSL TEAM picture

Welcome to medicalequipment-msl.com.

TECHNICAL DATA

Technical Specifications

Technical parameters from the currently available product information.

wavelengths
340~670nm
Throughput
Up to 100 tests/hour (without ISE), up to
Methodology
End-point, fxed-time, kinetic, optional
R1
180~450μl, step by 1μl
R2
30~250μl, step by 1μl
Sample volume
3~45μl, step by 0.5μl
Probe cleaning
Automatic washing both interior and exterior
dilution ratio up to 1
150
Reaction rotor
Rotating tray, containing 40 cuvettes
Cuvette
Optical length 5mm
Reaction volume
180~500μl
Reaction temperature
37±0.1?C
Mixing system
Independent mixing bar
Light Source
Halogen-tungsten lamp
Wavelength
340nm, 405nm, 450nm, 510nm
Linear range
0~3.5Abs
Calibration mode
Linear (one-point, two-point and multi-point)
Operation system
Windows? XP Professional/Home SP2, Windows? 7

Specifications can vary by configuration. Confirm intended use, final model, and supply details with our product team.

NEED A CONFIGURATION?

Discuss this product with our team

Share your application, destination, and required configuration for a project-specific response.

Contact Us