Lab Equipment > Rayto IVD

Rayto Chemray 420 auto chemistry analyzer for sale

MODEL / Chemray 420

Rayto Chemray 420 auto chemistry analyzer for sale. 360 tests/hour (single and dual reagent)

  • Automatic, Random Access
  • Automatic washing both interior and exterior, carryover <0.1%
PRODUCT OVERVIEW

Product Details

Quick Details

Throughput:

360 tests/hour (single and dual reagent)

Up to 540 tests/hour with ISE (K+, Na+, CI-)

Specifications

Technical Specifications

.System Function

Automatic, Random Access

STAT sample priority, available for all sample position

Barcode Scanner for sample and reagent

Bi-direction LIS interface (HL7)

Throughput:

360 tests/hour (single and dual reagent)

Up to 540 tests/hour with ISE (K+, Na+, CI-)

Principle: Colorimetric, Turbidimetric

Methodology:

End-point, Two-point, Kinetic Discrepant Value

Single/Dual/Triple/Quadruple reagent chemistries

Monochromatic/Bichromatic

Programming: Open system with user defined profiles and calculation assays

cbc medical Rayto Chemray 420

Probe cleaning:

Automatic washing both interior and exterior, carryover <0.1%

Pre-heating for reagent

Auto dilution:3-150 times

Built-in reagent/sample bar code reader (optional)

.Reaction System

Reaction positions: 100 cuvettes

Cuvette:Optical length 5mm

Reaction volume: 150.500u1

Reaction temperature:37'C,10.1℃

2 mixing probes

8 steps auto-washing system

.Sample/Reagent Handling

2 reagent probes, 1 sample probe

Sample tray:100 positions

Reagent tray: 80 refrigerated positions

Refrigeration temperature:2一100C

Reagent volume:20、350 NL, 1 pL adjustable

RI:150·350 pL, 1 pL adjustable

R2:20·250 NL,1 pL adjustable

Support R3 and R4

Sample volume: 2、45 pL, 0.1 pL adjustable

Sample and Reagent probe: Collision protection, liquid level detection and inventory checking

.Measuring and Optical System

Light Source:Halogen tungsten lamp

Photometer:Grating photometry

Wavelengths: 340nm,405nm, 450nm, 505nm, 540nm, 570nm, 600nm, 635nm,670nm,700nm,760nm,795nm

Absorption range:0~2.500Abs

Resolution:0.0001 Abs

.Working Condition

Power supply:100·240 VAC 50/60 Hz, 1500w

Temperature: 100C、300C

Humidity:40%~85%

Water consumption: s20 Uhour

Dimension: 1150mm(L)0726.5mm(W)01150mm(H)

.360 tests/hour, up to 540 tests/hour with ISE(optional)

.Grating system with 12 wavelengths

.Built-in bar code reader (optional)

.8 steps washing system

.Bi-direction LIS interface

TECHNICAL DATA

Technical Specifications

Technical parameters from the currently available product information.

Auto dilution
3-150 times
Reaction positions
100 cuvettes
Cuvette
Optical length 5mm
Reaction volume
150.500u1
Reaction temperature
37'C,10.1℃
reagent probes
1 sample probe
Sample tray
100 positions
Reagent tray
80 refrigerated positions
Refrigeration temperature
2一100C
Reagent volume
20、350 NL, 1 pL adjustable
RI
150·350 pL, 1 pL adjustable
R2
20·250 NL,1 pL adjustable
Sample volume
2、45 pL, 0.1 pL adjustable
Sample and Reagent probe
Collision protection, liquid level detection and inventory checking
Light Source
Halogen tungsten lamp
Wavelengths
340nm,405nm, 450nm, 505nm, 540nm, 570nm, 600nm, 635nm,670nm,700nm,760nm,795nm
Absorption range
0~2.500Abs
Resolution
0.0001 Abs

Specifications can vary by configuration. Confirm intended use, final model, and supply details with our product team.

NEED A CONFIGURATION?

Discuss this product with our team

Share your application, destination, and required configuration for a project-specific response.

Contact Us